Model Number CLV-190
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Event Description
The customer reported the manual brightness of the evis exera iii xenon light source would not get brighter.The issue was found during a diagnostic laparoscopic procedure that took an hour to complete.No error messages were displayed and just the view had dim light.The setup was noted to be normal and no other devices were connected besides the camera head and cord.The case was completed with a similar device.
 
Manufacturer Narrative
In speaking with olympus technical assistance center (tac) the customer tried unplugging and reseating both of the light cables to troubleshoot the issue.Reseating the cables resolved the issue and the customer confirmed that manual and auto brightness were working.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over six (6) years since the subject device was manufactured.Based on the results of the investigation, the root cause of the problem could not be identified, the probable cause: a human error made at the time the customer operated the light source.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17965706
MDR Text Key326063931
Report Number3002808148-2023-11488
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/19/2023
Supplement Dates Manufacturer Received11/13/2023
Supplement Dates FDA Received11/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-