• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G148
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system recorded a beeping alert.Upon interrogation of the device, an out of range pacing impedance measurement greater than 3000 ohms was observed for the right atrial (ra) lead.It was mentioned that in-office measurements show in-range impedance values around 500 ohms and no noise, however, when the patient is pushing downwards, the pacing impedance jumps to greater than 3000 ohms and artefacts are observed.Technical services (ts) recommended system revision to check lead integrity and the device header connection.The crt-d system remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system recorded a beeping alert.Upon interrogation of the device, an out of range pacing impedance measurement greater than 3000 ohms was observed for the right atrial (ra) lead.It was mentioned that in-office measurements show in-range impedance values around 500 ohms and no noise, however, when the patient is pushing downwards, the pacing impedance jumps to greater than 3000 ohms and artefacts are observed.Technical services (ts) recommended system revision to check lead integrity and the device header connection.The crt-d system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17967325
MDR Text Key326276070
Report Number2124215-2023-58173
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/28/2022
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number233275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/05/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-