MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105ELPKNA ELI LILY PINK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105ELPKNA ELI LILY PINK; SYRINGE, PISTON Back to Search Results
Model Number MMT-105ELPKNA
Device Problems Fluid/Blood Leak (1250); Appropriate Term/Code Not Available (3191)
Patient Problem Hyperglycemia (1905)
Event Date 09/26/2023
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer experienced hyperglycemia with a blood glucose value up to 500 mg/dl.Troubleshooting was performed and found that the customer did not change the needle after each injection.The cartridge and the cartridge holder were wet.No further patient complications were reported.The customer will continue the use of the insulin pen and will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELPKNA ELI LILY PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key17971958
MDR Text Key326124424
Report Number3012822846-2023-01724
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000337
UDI-Public(01)000010862088000337(17)250516
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-105ELPKNA
Device Catalogue NumberMMT-105ELPKNA
Device Lot NumberB1781
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/27/2023
Initial Date FDA Received10/20/2023
Date Device Manufactured05/16/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age84 YR
Patient SexFemale
Patient Weight55 KG
-
-