Reportable based on device analysis completed on 18oct2023.It was reported that balloon inflation failure occurred.The 70% stenosed target lesion was located in the moderately tortuous and moderately calcified left arm.A 6.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during procedure, the balloon could not be inflated and after the removal of the device from the patient, it was found that the balloon could no longer be used.The device was simply removed, and the procedure was completed with another of the same device.No patient complications were reported, and the patient was stable.However, returned device analysis revealed balloon longitudinal and circumferential tear.
|