Catalog Number 8608500 |
Device Problems
Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the ventilator failed during use.There was no patient injury reported.
|
|
Manufacturer Narrative
|
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
|
|
Event Description
|
It was reported that the ventilator failed during use.There was no patient injury reported.
|
|
Manufacturer Narrative
|
At the time reported, the log file showed a sporadic failure of the inspiratory pressure sensor (pz) on the vgc analog circuit board.The device reacted as specified with an autonomous shutdown of the ventilator and gave an acoustic and visual alarm for ventilator fail, while switching autonomously to manual ventilation (man/spont).Gas dosing and monitoring are not affected.
|
|