ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Catalog Number 00882100100 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined no jamming was noted; however, the motor was corroded and ran below specification.The motor was replaced to resolve the issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.G2, country event occurred in: (b)(6).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during testing prior to surgery, that the unit was jamming frequently.There was no reported harm, injury, or delay as there was no patient involvement.Due diligence is complete.No additional information is available.After evaluation of the returned device, it was observed that the device had inconsistent speed.No adverse event reported.
|
|