(b)(4).The complaint sample was returned.Tecomet, the manufacturing site reported: "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, inc.Kenosha, wi facility as part of a 50-pc.Lot in february of 2023.It was found that this order was made from the correct materials and components and all approved processes were followed.Evaluation of the returned instrument shows that the jaws are loose and misaligned to each other and bent to the left and the jaw pivot pin is pulled thru one side of the damaged/bent outer tube assembly.(pie's attached) we are able to validate this complaint for the returned instrument.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) is bent, and its bosses are both damaged where they engage the jaws.We suspect that the damaged drive rod bosses caused the jaws to become slightly loose and misaligned in the closed position.We are unable to determine what caused all the damage to the components on this device but mishandling such as excessive force being applied to the handle to jaw mechanism on this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.Teleflex will continue to monitor and trend for reports of this nature.
|