MAUDE Adverse Event Report: QUIDELORTHO CORPORATION SOFIA2 FLU + SARS ANTIGEN FIA, NS SWAB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUIDELORTHO CORPORATION SOFIA2 FLU + SARS ANTIGEN FIA, NS SWAB Back to Search Results
Catalog Number 20377
Device Problems No Apparent Adverse Event (3189); Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2023
Event Type  malfunction  
Event Description
Customer reporting that a new employee during training accidently swabbed another employee in the nose using a positive control swab (swab is coated with non-infectious recombinant flu a, flu b, and sars antigens).Employee was immediately tested for sofia flu+sars and tested negative for all targets (flu a, b, sars).Employee has not experienced any symptoms or adverse reactions to date.
 
Manufacturer Narrative
Investigation summary: a review of the package insert (pi) was conducted for clarity of instructions.No issues were found.The customer's reported problem was related to a deviation from the instructions called out in the pi.Root cause: customer procedural error.Source: phone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFIA2 FLU + SARS ANTIGEN FIA, NS SWAB
Type of Device
SOFIA2 FLU + SARS ANTIGEN FIA, NS SWAB
Manufacturer (Section D)
QUIDELORTHO CORPORATION
10165 mckellar court
san diego CA 92121
Manufacturer Contact
karl luke
10165 mckellar court
san diego, CA 92121
MDR Report Key17981277
MDR Text Key326358075
Report Number0002024674-2023-01801
Device Sequence Number1
Product Code QMN
UDI-Device Identifier30014613339229
UDI-Public30014613339229
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EUA202751
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number20377
Device Lot Number708405
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-