• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND CO. BD NANO¿ 2ND GEN PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND CO. BD NANO¿ 2ND GEN PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Catalog Number 320550
Device Problem Defective Component (2292)
Patient Problem Easy Bruising (4558)
Event Date 10/16/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd nano¿ 2nd gen pen needle was difficult to operate during use.The following information was provided by the initial reporter: "consumer reported that the patient end needle bends during injection.Consumer also reported that the pen jams during injection.Also reported that the needle slipped once as she was injecting and bruised her skin.Consumer does not re-use.".
 
Event Description
It was reported that the bd nano¿ 2nd gen pen needle was difficult to operate during use.The following information was provided by the initial reporter: "consumer reported that the patient end needle bends during injection.Consumer also reported that the pen jams during injection.Also reported that the needle slipped once as she was injecting and bruised her skin.Consumer does not re-use.".
 
Manufacturer Narrative
The following fields have been updated due to additional information: h.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NANO¿ 2ND GEN PEN NEEDLE
Type of Device
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer (Section G)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17981576
MDR Text Key326388661
Report Number9616656-2023-01128
Device Sequence Number1
Product Code FMI
UDI-Device Identifier00382903205509
UDI-Public00382903205509
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K212015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number320550
Device Lot Number2320083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-