Model Number CLH-250
Device Problems Failure to Power Up (1476); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found another reportable malfunction reported in the b5 section.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus that during reprocessing, the halogen light source does not start.There was no procedure involved and no reports of patient or user harm.The device was returned for evaluation.During the device evaluation, another reportable malfunction was found: fan failure.This medical device report (mdr) is being submitted to capture both reportable malfunctions: one reported by the customer and the other found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to a defective lamp socket and defective fan.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17982169
MDR Text Key326273332
Report Number3002808148-2023-11580
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170024429
UDI-Public04953170024429
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K981543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLH-250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-