Model Number ZINC AIR BATTERIES |
Device Problems
Fluid/Blood Leak (1250); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/25/2023 |
Event Type
malfunction
|
Event Description
|
The user claimed that the battery burst open while using the audio processor.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Manufacturer Narrative
|
Conclusion: according to the information received the battery burst open.The device investigation revealed a separated zinc air battery with leakage.Also contamination of the battery pack cover could be observed.In previous cases seen by the supplier, the cell expansion happened due to an abnormal use, e.G., cell is being charged or repeated use of empty, completely discharged cell.The recipient did not get injured due to this issue.This is a final report.
|
|
Event Description
|
The user claimed that the battery burst open while using the audio processor.The audio processor is still functional.No injury was reported.
|
|