ENTERRA MEDICAL, INC. ENTERRA II IPG; GASTRIC STIMULATOR Back to Search Results
Model Number 37800
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 06/01/2023
Event Type  malfunction  
Event Description
Patient received a new implant in december.First week was great for her, then felt shocking sensation.I down adjusted voltage but she was always able to feel this.We replaced the battery in january, she did fine with this for 3 months.Recently reported return of shocking sensation, all impedance and lead analysis are ok and within normal parameters.She reports the stimulation therapy improves her symptoms, and would like to keep the device.Dr.Friedman and i think maybe we need to replace the entire unit? prior to explanting for good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA II IPG
Type of Device
GASTRIC STIMULATOR
Manufacturer (Section D)
ENTERRA MEDICAL, INC.
5353 wayzata blvd ste. 400
saint louis park 55416
MDR Report Key17986508
MDR Text Key326310482
Report Number3027386225-2023-00003
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37800
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age41 YR
Patient SexFemale
-
-