Model Number 8300AB25
Device Problems Gradient Increase (1270); Insufficient Information (3190)
Patient Problems Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Event Date 09/18/2023
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was learned via the patient registry that a patient with a 25mm 8300ab intuity valve underwent a valve-in-valve procedure after an implant duration of five (5) years, 11 months due to unknown reasons.The tavr procedure was performed with a 23mm 9755rsl transcatheter valve.The patient was noted to be in recovery post procedure.
 
Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bio-prosthesis explant and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A capa/scar/pra is not required as there are no confirmed product or labeling non-conformances and no other triggers are met.The most likely cause is patient factors, including hyperlipidemia and diabetes mellitus.
 
Event Description
It was learned via the patient registry that a patient with a 25mm 8300ab intuity valve underwent a valve-in-valve procedure after an implant duration of five (5) years, 11 months due to aortic stenosis.The patient initially presented with heart failure and shortness of breath.The tavr procedure was performed with a 23mm 9755rsl transcatheter valve.The patient was noted to be stable in recovery post procedure.The patient was discharged on pod #1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key17986881
MDR Text Key326295723
Report Number2015691-2023-16989
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date12/08/2018
Device Model Number8300AB25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2023
Date Device Manufactured12/08/2016
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization; Required Intervention;
Patient Age62 YR
Patient SexMale
-
-