Model Number A22001A |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The customer reported to olympus that the telescope, 12°, 4 mm had the guide pin bending, cutting.The event occurred during cleaning at the end of the operation.The procedure was completed with the same set of equipment.There was no patient harm associated with the event.
|
|
Manufacturer Narrative
|
The device was returned and evaluated, and the customer¿s allegation was confirmed.Additional findings were the outer tube was bent and there was deterioration of the eyepiece ring.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.E1: (b)(6).
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling and application of excessive force.Olympus will continue to monitor field performance for this device.
|
|