• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 12°, 4 MM; OLYMPUS HYSTERORESECTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 12°, 4 MM; OLYMPUS HYSTERORESECTOSCOPE Back to Search Results
Model Number A22001A
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the telescope, 12°, 4 mm had the guide pin bending, cutting.The event occurred during cleaning at the end of the operation.The procedure was completed with the same set of equipment.There was no patient harm associated with the event.
 
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was confirmed.Additional findings were the outer tube was bent and there was deterioration of the eyepiece ring.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.E1: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling and application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 12°, 4 MM
Type of Device
OLYMPUS HYSTERORESECTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17994421
MDR Text Key326359732
Report Number9610773-2023-03045
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020848
UDI-Public04042761020848
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22001A
Device Lot Number545404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-