Catalog Number 762165 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to damaged product as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode damaged product.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that after collection 18 paxgene® blood rna tube the tubes cracked when placed in freezer.There was no report of patient impact.The following information was provided by the initial reporter: 18 samples collected into paxgene rna tubes 762165, tubes were placed directly into -70 c freezer, and all 18 tubes cracked.
|
|
Search Alerts/Recalls
|
|
|