• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS Back to Search Results
Catalog Number 762165
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to damaged product as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode damaged product.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that after collection 18 paxgene® blood rna tube the tubes cracked when placed in freezer.There was no report of patient impact.The following information was provided by the initial reporter: 18 samples collected into paxgene rna tubes 762165, tubes were placed directly into -70 c freezer, and all 18 tubes cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17999437
MDR Text Key326569586
Report Number9617032-2023-01540
Device Sequence Number1
Product Code NTW
UDI-Device Identifier54053727621653
UDI-Public(01)54053727621653
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082150
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number762165
Device Lot Number2144440
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-