B3-date of event is estimated.Additional components potentially involved in the event: common device name: drg: lead, model: mn20450-50a, udi: (b)(4), serial: (b)(6), batch: 8830806.Common device name: drg: lead, model: mn20450-50a, udi: (b)(4), serial: (b)(6), batch: 8892605.
|
The report event of slim tip broken was not confirmed.As received, microscopic inspection showed the returned lead slim tip no anomalies/damage was found.The cause of the reported event remains unknown.The report event of high impedances was not confirmed.As received, electrically tested showed the returned both terminal ends segments were passed isolation resistant testing.The cause of the reported event remains unknown.".
|