Model Number UI500
Device Problems No Display/Image (1183); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2022
Event Type  malfunction  
Event Description
It was reported that the device stopped working during a procedure and needed to be replaced by a back-up unit.The patient was not harmed but bringing in a re-placement for the device resulted in a delay of the procedure.
 
Manufacturer Narrative
The previous reportability decision for this event was reversed based on a retrospective review.Primary damage location: touch-screen.Primary damage type: short-circuit.Primary damage cause: wear/often used.Details: touch screen short circuit, filter board and sensor interface board upgrade required.Pressure sensor board failure error 322.Arrived with a broken temp sensor cap on faceplate.Internal karl storz reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOFLATOR 50
Type of Device
ENDOFLATOR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key18000437
MDR Text Key326614737
Report Number9610617-2023-00327
Device Sequence Number1
Product Code FCX
UDI-Device Identifier04048551326657
UDI-Public4048551326657
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUI500
Device Catalogue NumberUI500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-