A.1.-a.5.There was no known patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in china.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
All functional tests positively passed and the unit was put back into service.Complaints database analysis revealed no similar event since unit installation in 2021.Based on all the above facts and the results of a similar complaints database analysis, it cannot be ruled out that the most likely root cause of the reported event is a non-free to spin pump motor, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, that required a power overload to work, which could have resulted in an over-current ultimately caused damage to involved board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|