Model Number 16-02-85 |
Device Problems
Fluid/Blood Leak (1250); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6) minnesota.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, leaking sniffle valve was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t was leaking a large amount of water from the patient 1 circuit on the input side at the venting valve.The issue occurred at setup.There was no patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
Venting valve was replaced two weeks before during pm.A device service history review has been performed and identified that the unit was manufactured in 2015 and no other similar event has been reported, neither concerning trend has been identified.Based on the information provided by field service technician, the most likely root cause of the reported issue is an adhesive failure of the recently replaced venting valve.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|