Model Number N/A |
Device Problems
Corroded (1131); Unstable (1667); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was sent in for preventative maintenance but repair was needed.The motor speed was unstable; there were worn screws; and the motor was corroded.There was no harm, injury, or delay as there was no patient involvement.Due diligence is complete.
|
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4).G2, country event occurred in: poland.The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
|
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4) review of the most recent repair record determined the motor speeds were unstable, the motor was corroded, and the screws were worn.The motor and screws were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information is available regarding the incident.
|
|