Model Number 1258T/86
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 10/10/2023
Event Type  Injury  
Event Description
It was reported that during initial implant procedure, patient's new left ventricular (lv) lead was not able to be implanted due to patient anatomy.It was further reported that patient experienced heart failure.The lead was not installed and the procedure was completed using alternate lead on (b)(6) 2023.The patient was stable after procedure.
 
Manufacturer Narrative
The reported event was of the surgical complications.As received, a complete lead was returned in one piece.Visual examination of the lead did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.S-curve was measured to be within specification.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18006710
MDR Text Key326530812
Report Number2017865-2023-50654
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1258T/86
Device Lot NumberA000136038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-