Model Number HL-90
Device Problems Electrical /Electronic Property Problem (1198); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Event Description
It was reported that the device electrical safety failed.Patient involvement unknown.
 
Manufacturer Narrative
Other, other text: d4: udi number is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
Additional information was received: event occurred during preventive maintenance.No medical intervention was required but is still ongoing.Relevant tests or laboratory data: "multiple tests run with esa, ground wire resistance was consistently above 0.5 ohms, which is our standard".
 
Manufacturer Narrative
Other, other text: additional information: a1.B5 and h6 - health effects codes.
 
Manufacturer Narrative
Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Email is: (b)(4).One device was received.Per visual inspection, old, faded line cord, out dated front cover, out dated printed circuit board (pcb) and power supply, corroded quick connect, corroded thermistor, enclosure had multiple nicks and scratches, water tank cover was cracked on all four corners, inside water tank was rust contamination.Per functional testing, the device did not pass electrical safety.The complaint was confirmed.The root cause was a bent micro switch, it was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18007587
MDR Text Key326848678
Report Number3012307300-2023-09917
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received12/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-