The manufacturer became aware of an allegation that a simplygo device had a failure description of a triangle with arch to the right.There was no report of patient harm or injury.During the evaluation of the device at a third-party service center was observed low o2, sieve needs to be replaced, muffler kit, both pressure valve and tubing are contaminated with black dust, inlet filter needs to be changed due to preventive maintenance, compressor does not start.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
|