MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXUSA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction drive tube leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot m315 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot m315 shows no trends.Trends were reviewed for complaint category drive tube leak/break.No trends were detected for this complaint category.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer contacted mallinckrodt to report they experienced a drive tube leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported the drive tube broke after approximately 1500 ml of whole blood had been processed.The customer aborted the ecp treatment and no residual blood within the kit was returned to the patient.The customer reported the patient was in stable condition.The kit return was requested; however, the customer had discarded the kit.No product was returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin, D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS, INC.
2201 bailey ave
buffalo NY 14211
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key18011891
MDR Text Key326620826
Report Number3013428851-2023-00064
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)M315(17)250301
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCLXUSA
Device Lot NumberM315
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/29/2023
Initial Date FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient Weight70 KG
-
-