• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6201A
Device Problems Loss of Data (2903); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the central nurse's station (cns) is having issues with waveforms freezing and disappearing, but the heart rate remains still, this happens about every 8 hours.The bme checked the logs and noticed an error message saying, "mcafee solidifier prevent unauthorized execution of windows".No patient harm was reported.Technical support recommended that the bme send this unit into nk repair center.Service performed: device was received into nihon kohden repair center and inspection found there was no physical damage or fluid intrusions.The reported problem of waveforms freezing and disappearing, was duplicated.The system board is corrupted.There were beeping noises when turning on the unit.The motherboard pcb is no longer being supported by nkc.It is recommended that the customer do an exchange for this unit.This unit has been in use since 2017 and review of the service history for this sn cns shows that it has never been sent into nka for a hard drive replacement.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The biomedical engineer (bme) reported that the central nurse's station (cns) is having issues with waveforms freezing and disappearing, but the heart rate remains still, this happens about every 8 hours.The bme checked the logs and noticed an error message saying, "mcafee solidifier prevent unauthorized execution of windows".No patient harm was reported.Technical support recommended that the bme send this unit into nk repair center.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18014363
MDR Text Key327299651
Report Number8030229-2023-03834
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public04931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-