The biomedical engineer (bme) reported that the central nurse's station (cns) is having issues with waveforms freezing and disappearing, but the heart rate remains still, this happens about every 8 hours.The bme checked the logs and noticed an error message saying, "mcafee solidifier prevent unauthorized execution of windows".No patient harm was reported.Technical support recommended that the bme send this unit into nk repair center.Service performed: device was received into nihon kohden repair center and inspection found there was no physical damage or fluid intrusions.The reported problem of waveforms freezing and disappearing, was duplicated.The system board is corrupted.There were beeping noises when turning on the unit.The motherboard pcb is no longer being supported by nkc.It is recommended that the customer do an exchange for this unit.This unit has been in use since 2017 and review of the service history for this sn cns shows that it has never been sent into nka for a hard drive replacement.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
The biomedical engineer (bme) reported that the central nurse's station (cns) is having issues with waveforms freezing and disappearing, but the heart rate remains still, this happens about every 8 hours.The bme checked the logs and noticed an error message saying, "mcafee solidifier prevent unauthorized execution of windows".No patient harm was reported.Technical support recommended that the bme send this unit into nk repair center.
|