Model Number CI422
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/30/2023
Event Type  Injury  
Manufacturer Narrative
This report is submitted on october 27, 2023.
 
Event Description
Per the clinic, the patient experienced an incorrect placement and poor sound with the device.Subsequently, the device was explanted on (b)(6) 2023.There are no plans to re-implant the patient with a new device as of the date of this report.
 
Manufacturer Narrative
Per the clinic, the patient experienced a performance decrement due to migration of the electrode array.This report is submitted on november 21, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI422 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
aini yusof
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18017217
MDR Text Key326673861
Report Number6000034-2023-03455
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502020268
UDI-Public(01)09321502020268(11)130422(17)150421
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/21/2015
Device Model NumberCI422
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-