• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Unstable (1667)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The device was sent in for preventative maintenance originally and found to be needing repaired for motor speed unstable and corroded motor.There was no alleged event or malfunction reported for this device when it was sent in for maintenance.Follow-up will not be conducted.Due diligence is complete.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: united kingdom.Once the investigation is complete, a follow up/final report will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the corroded motor had unstable speed.The motor was replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18021151
MDR Text Key326724305
Report Number0001526350-2023-01383
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375994
UDI-Public(01)00889024375994(11)190123(10)64277079
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64277079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-