Model Number 3662
Device Problem Disconnection (1171)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/04/2023
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that the patient's lead has pulled out of the ipg.Surgical intervention may be pending to address the issue.
 
Manufacturer Narrative
The reported observation of lead disconnecting from the ipg was not confirmed.The root cause of the reported observation was not ascertained.The lead in question was not returned for analysis.Using clinicians programmer, ipg system impedance testing was within the expected range.The ipg was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.The ate test specifically tests the ipg output and system impedances on all 16 electrodes including the ipg can connection.A lead pull test confirmed the lead remained fully seated and properly engaged with the ipg setscrews.
 
Event Description
It was reported patient underwent surgical intervention on (b)(6) 2023 during which ipg was explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18021690
MDR Text Key326736806
Report Number3006705815-2023-06943
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031426
UDI-Public05415067031426
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number3662
Device Lot NumberA000126725
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD
Patient Outcome(s) Other;
Patient Age37 YR
Patient SexFemale
Patient Weight116 KG
-
-