• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS 320
Device Problems Failure to Interrogate (1332); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted product was not returned to neuropace for analysis.
 
Event Description
The patient had a resection on (b)(6) 2023.During the resection, electrocautery was used near one of the leads.The rns was not able to be interrogated following the resection procedure.On (b)(6) 2023, the rns was replaced due to electrocautery damage.Neuropace was not able to acquire the rns that was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
5108822607
MDR Report Key18022825
MDR Text Key326749187
Report Number3004426659-2023-00055
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005366
UDI-Public010085554700536617210827
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberRNS 320
Device Catalogue Number1007927
Device Lot Number30524-1-1-1
Date Manufacturer Received10/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age57 YR
Patient SexFemale
-
-