Model Number A22041A
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported that the cap of the sheath was detached.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: scraping on the inside of the tip of the beak.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported issue was confirmed, the cap was detached.The device evaluation found scraping on the inside of the tip of the beak.It was also found that the rubber seal was discolored and there was damage on the tension ring.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 15 years since the subject device was manufactured.Based on the results of the investigation, root cause of the reported damage is consistent with wear from use over time and improper handling.However, specific root cause could not be determined.The issue can be detected/prevented by following the instructions provided in: the instructions for use, chapter 4 before use ¿ warning infection control risk - properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing - inspecting the product - visually inspect the product.Make sure that it has: no corrosion no dents no scratches ceramic insulation at distal end - visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning - risk of injury - impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product - if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR.
Type of Device
RESECTION SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18023241
MDR Text Key326753181
Report Number9610773-2023-03099
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020961
UDI-Public04042761020961
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22041A
Device Lot Number08XW-0025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-