Model Number CLV-190
Device Problems Communication or Transmission Problem (2896); Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Event Description
The customer reported to olympus that during an inspection for use, the evis exera iii xenon light source had no image, displayed errors b30 and e216, and presumably had moisture in the socket.There were no reports of patient/user harm.
 
Manufacturer Narrative
The device was evaluated by an olympus technician onsite, the customer's allegation was not confirmed.The technician confirmed that the light source had no problem, but the endoscopes were all full of water.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Manufacturer Narrative
Corrected data: d9 (¿no¿ was selected in error on the initial report), h3 (¿no¿ and ¿not returned to manufacturer¿ were selected in error on the initial report), and h6 (type of investigation code ¿4114¿ was selected in error on the initial report instead of ¿10¿) this report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the b30 error and water penetrating into the power plug of the device could not be identified.However, it¿s likely the cause of the b30 error is related to a scope communication error.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18025213
MDR Text Key326816905
Report Number3002808148-2023-11944
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-