• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APIFIX LTD. MID-C 115; POSTERIOR RATCHETING ROD SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APIFIX LTD. MID-C 115; POSTERIOR RATCHETING ROD SYSTEM Back to Search Results
Model Number MID-C 115
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Manufacturer Narrative
Production process analysis: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.User (surgeon and patient) information analysis: on 02-oct-2023 apifix was notified that patient (b)(6) is scheduled for revision surgery on (b)(6) 2023 due to implant breakage (rod).The plan is to replace to a new apifix device.On (b)(6) 2023 apifix was notified that the revision surgery was performed on that day, during which the broken implant (mid-c 115) was replaced with a new one (mid-c 125). the broken implant is expected to be returned to the manufacturer for analysis.Apifix clinical affairs reviewed the patient case file and identified that patient (b)(6) was treated out of indications for use.This patient had a cobb anlge of 57° with lateral bending 32° (mid-c ifu  - apifix is indicated for use in patients having a cobb angle of up to 60 degrees which reduces to less than or equal to 30 degrees on lateral side-bending radiographs) risk assessment: implant breakage can result from trauma, practicing contact or high demand sports, hyper-kyphosis, inserting the pedicle screws in a wrong trajectory which locks the poly-axial joints, not working according to the surgical technique (e.G., minimally invasive approach and/or not using the trial tool properly).Re-operation events are a known risk that was assessed and recorded by the product risk assessment.The risk of mechanical failure of the mid-c mechanism - [misuse > implant breakage] has been assessed and found to be acceptable.The implant breakage rate due to any reason is in line with the rate reported in the literature for this type of complication as described in the company's clinical evaluation report.The explanted device is expected to be returned to the manufacturer for analysis.Upon completion of the evaluations, when additional information comes to light, then a supplemental medwatch report will be submitted.
 
Event Description
On 02-oct-2023 apifix was notified that patient (b)(6) is scheduled for revision surgery on (b)(6) 2023 due to implant breakage (rod).The plan is to replace to a new apifix device.
 
Manufacturer Narrative
Return analysis: the explanted device was returned and was subjected to cleaning, steam sterilizing, and engineering evaluation.The device was obviously fractured in the pole component around the mid-point of the pole.It appears that post fracture the device remained implanted as the fracture plane, body, and side of pole showed signs of significant wear.The fracture plane was worn completely smooth from and thus fracture mode is not able to be identified.No wear was visibly observed on the spherical ring of the pole.Heavy wear was observed on the spherical ring of the base.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 115
Type of Device
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel st.
yokneam ellit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key18027617
MDR Text Key326804101
Report Number3013461531-2023-00050
Device Sequence Number1
Product Code QGP
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
H170001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMID-C 115
Device Catalogue NumberAFS-115-040
Device Lot NumberAF-11-01-18
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age20 YR
Patient SexMale
-
-