AIZU OLYMPUS CO., LTD. EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H260Z
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegations was confirmed.The evaluation found that due to a pinhole on air/water tube, water tightness is lost; and due to damage on coupled charging device (ccd) unit, the image is not displayed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus the evis lucera gastrointestinal videoscope, had a blurry image.The event was found during reprocessing.The procedure was unknown.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
A review of the device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, the air water (a/w) channel leaking during usage led to humidity entering the objective (ob) lens unit resulting in the image becoming blurred.The event can be averted by following the instructions for use (ifu) sections which state: "chapter 3 preparation and inspection before each case, prepare and inspect this instrument as instructed below.Inspect other equipment to be used with this instrument as instructed in their respective instruction manuals.Should any irregularity be observed after inspection, follow the instructions as described in chapter 5, ¿troubleshooting¿.If this instrument malfunctions, do not use it.Return it to olympus for repair as described in section 5.3, ¿returning the endoscope for repair¿ on page 99.Warning using an endoscope that is not functioning properly may compromise patient or operator safety and may result in more severe equipment damage." "chapter 5 troubleshooting if the endoscope is visibly damaged, does not function as expected, or is found to have irregularities during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope.Contact olympus.Some problems that appear to be malfunctions may be correctable by referring to section 5.1, ¿troubleshooting guide¿.If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to olympus for repair.Warning never use the endoscope on a patient if an irregularity is observed.Damage or an irregularity in the instrument may compromise patient or user safety and may result in more severe equipment damage.Should any irregularity in the function of the endoscope and/or endoscopic image be observed during use, stop the examination immediately and carefully withdraw the endoscope from the patient as described in section 5.2, ¿withdrawal of the endoscope with an irregularity¿ on page 97." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18030866
MDR Text Key326865055
Report Number9610595-2023-16093
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H260Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-