An unexpected application error issue was reported with the adc application version 2.10.0, in use with iphone 7 with ios operating system version 1577.Adc has identified that following the release of the freestyle librelink (fsll) application for ios, v2.10.0, on (b)(6) 2023 (11am) in the united kingdom (uk), some users have experienced a situation where the application upgrade is unsuccessful, and as a result, they receive a white screen in the application user interface (ui).This issue does not affect users in the united states.This will result in a period where glucose readings and alarms will not be received, and historical glucose readings will not be accessible, as the previous fsll application, v2.8.1, was no longer available to users on the apple app store.This issue was addressed in the field by adc fa1029-2023 and was resolved in the field by the release of updated fsll ios application, made available in the apple app store in the uk on (b)(6) 2023.There was no report of death, serious injury, or mistreatment associated with this event.This report is being submitted as part of retrospective reporting deemed necessary related to fa1029-2023.
|
This complaint was investigated.Adc has identified a software defect for the freestyle librelink application for ios, version 2.10.0, in which the application upgrade was unsuccessful.This resulted in a period where users may not have received glucose results or may not have been alerted to low or high glucose alarms.Based on the investigation, the complaint is confirmed and no further investigation activities for this individual complaint are required.This issue was addressed in the field by adc fa1029-2023.The device model number populated in section d4 is for the freestyle librelink ios application, on market in the uk.This is same/similar to us freestyle librelink/freestyle libre 2 ios application part number 71733-01/71926-01.All pertinent information available to abbott diabetes care has been submitted.
|