It was reported to boston scientific corporation that a polaris ultra ureteral stent was used during a ureteroscopic lithotripsy procedure in the ureter performed on (b)(6)2023.During the procedure, when the device was unpacked, it was found that the coil was partially broken.It was noted that the coil was detached from the stent.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.This event has been deemed a reportable event based on the investigation finding of stent, detached/ separated please see block h10 for full investigation details.
|
Block h6: imdrf device code a0401 captures the reportable investigation results of stent shaft detached.Block h10: the returned polaris ultra ureteral stent was analyzed, and a visual evaluation noted that the shaft was detached.Additionally, the suture was not returned.During magnification, it was observed closely that the shaft was detached.No other problems with the device were noted.The reported event of stent coil detached was not confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.Based on the analysis performed, it is possible to conclude that this problem could be caused by operational factors.The retrieval line may be removed before placement at the physician's discretion.If the retrieval line was removed using excess force, the stent can get detached during the preparation, affecting the performance of the device.Consequently, affects the performance of the device.Therefore, the most probable root cause is adverse event related to the procedure.
|