BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL Back to Search Results
Model Number M0061921330
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a polaris ultra ureteral stent was used during a ureteroscopic lithotripsy procedure in the ureter performed on (b)(6)2023.During the procedure, when the device was unpacked, it was found that the coil was partially broken.It was noted that the coil was detached from the stent.The procedure was successfully completed with another polaris ultra ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.This event has been deemed a reportable event based on the investigation finding of stent, detached/ separated please see block h10 for full investigation details.
 
Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable investigation results of stent shaft detached.Block h10: the returned polaris ultra ureteral stent was analyzed, and a visual evaluation noted that the shaft was detached.Additionally, the suture was not returned.During magnification, it was observed closely that the shaft was detached.No other problems with the device were noted.The reported event of stent coil detached was not confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.Based on the analysis performed, it is possible to conclude that this problem could be caused by operational factors.The retrieval line may be removed before placement at the physician's discretion.If the retrieval line was removed using excess force, the stent can get detached during the preparation, affecting the performance of the device.Consequently, affects the performance of the device.Therefore, the most probable root cause is adverse event related to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS ULTRA
Type of Device
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086834015
MDR Report Key18038815
MDR Text Key326964660
Report Number3005099803-2023-05717
Device Sequence Number1
Product Code FAD
UDI-Device Identifier08714729124443
UDI-Public08714729124443
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K010002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2024
Device Model NumberM0061921330
Device Catalogue Number192-133
Device Lot Number0026851189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-