Catalog Number 04404483190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/14/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cobas 6000 c 501 serial number was (b)(6).The investigation is ongoing.
|
|
Event Description
|
There was an allegation of a questionable gluc hk gen.3 (glucose) result from the cobas 6000 c 501.The initial result was 2 mg/dl and was reported outside of the laboratory.The repeat result was 77 mg/dl and was believed correct.
|
|
Manufacturer Narrative
|
The customer canceled a scheduled service visit and stated they resolved the issue by recalibrating the assay.The investigation determined the customer's actions resolved the issue.
|
|