Model Number MR290V |
Device Problem
Overfill (2404)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The subject mr290v vented autofeed humidification chamber has been requested to be returned to fisher & paykel healthcare for evaluation.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A distributor reported on behalf of a healthcare facility in japan, via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber was overfilled before patient use.There was no patient involvement.
|
|
Event Description
|
A distributor reported on behalf of a healthcare facility in (b)(6), via a fisher and paykel healthcare (f&p) field representative, that a mr290v vented autofeed humidification chamber overfilled before patient use.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).Method: the subject mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand for investigation, where it was visually inspected and performance tested.Results: visual inspection of the returned chamber did not identify any damage or defects.Performance testing confirmed that the chamber filled with water and stopped when the maximum water line was reached.Conclusion: we are unable to determine the cause of the reported event as no fault was found with the returned device.The mr290v vented autofeed humidification chamber is designed with both a primary and secondary float system.If there is a malfunction of the primary float system the water may rise above the maximum water level line however the secondary float system is designed to prevent any further overfilling.Mr290v vented autofeed humidification chambers are functionally tested during production.Any chamber that fails is rejected.The subject mr290v chamber would have met the required specification at the time of production.The user instructions that accompany the rt380 adult dual-heated evaqua2 breathing circuit specifies in the warning section "do not use the chamber if the water rises above the maximum water level line" along with a diagram directing the user to check the water level on the chamber.A written description of the meaning of the symbols used is also included in the user instructions.It also states the following: "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.".
|
|
Search Alerts/Recalls
|