(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions for use (ifu) provided with this kit warns the user, "use only lumen(s) labelled "pressure injectable" for pressure injection to reduce risk of catheter failure and/or patient complications.Refer to the arrow pressure injection information label for pressure injection information." a finding in regard to extension line juncture hub separation in use was identified upon device history record review for which further investigations were initiated, but it could not be determined if this is relevant to this reported customer issue without the device sample to evaluate.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|