Catalog Number D139401
Device Problems Signal Artifact/Noise (1036); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an unspecified ablation procedure with a qdot-micro, uni-directional, d curve, c3, split handle and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, noise was detected on intracardial signals (map port 1-2) of the qdot catheter.After replacing of the appropriate catheter cable the problem with this catheter was not solved.This catheter was replaced with another one and from the same type, so that procedure was ended successfully.So there was no damage and consequences caused to the patient.The noise issue is not mdr-reportable.The hole in the pebax is mdr-reportable.
 
Manufacturer Narrative
E1 initial reporter phone: (b)(6).The photo analysis and product investigation were completed.Device evaluation details: a picture was received for evaluation following biosense webster's procedures.According to pictures provided by the customer, on the printing of the cardiolab screen was observed a signal noise.Additionally, the product was returned to biosense webster for evaluation.Bwi conducted a visual inspection and functional test of the returned device.Visual inspection was performed, and a hole was observed on the pebax surface with reddish material inside.The root cause of the damage on pebax cannot be related to the manufacturing process since there is evidence that the device was manufactured in accordance with documented specifications and procedures.However, it could be related to the handling of the device during the procedure, but this cannot be conclusively determined.An electrical test was performed, and no electrical issues were found.The blood inside the pebax could be related to the electrical issue reported by the customer.The customer complaint was confirmed.A manufacturing record evaluation was performed for the finished device batch number 31049025l, and no internal actions were identified.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT-MICRO, UNI-DIRECTIONAL, D CURVE, C3, SPLIT HANDLE
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18048287
MDR Text Key327628838
Report Number2029046-2023-02484
Device Sequence Number1
Product Code OAE
UDI-Device Identifier10846835017045
UDI-Public10846835017045
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139401
Device Lot Number31049025L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2023
Date Manufacturer Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-