Model Number PRT-01185-002
Device Problems Connection Problem (2900); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Event Description
It was reported that the device needed service.Upon evaluation of the device, ventec observed that the ventilator was unable to maintain peep (positive end expiratory pressure) and that it provided a patient circuit disconnect alarm.There was no patient involvement associated with the reported event.
 
Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of it being unable to maintain peep (positive end expiratory pressure) and alarming patient circuit disconnect was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issues.Proper device operation was then observed through functional and performance testing.The investigation determined that the cause of the reported issues was the lift.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key18048670
MDR Text Key327281611
Report Number3013095415-2023-00683
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00850018761154
UDI-Public0100850018761154
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPRT-01185-002
Device Catalogue NumberPRT-01185-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2022
Date Manufacturer Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-