ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71953-01
Device Problem Computer Software Problem (1112)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 10/19/2023
Event Type  Injury  
Event Description
Customer reported her adc blood glucose meter was reading in the incorrect uom and incorrect language.Customer further reported experiencing a loss of consciousness.Customer reported she was unable to self-treated requiring third-party treatment from a healthcare professional of "dextro" and something to eat.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The most probable root causes associated with this failure mode are software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the freestyle libre reader were reviewed and the dhrs showed the freestyle libre reader passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
Reader (b)(6)has been returned and investigated.Visual inspection has been performed on the reader and no issues were observed.The unit of measurement were set successfully.The returned reader was connected to pc and investigation was performed for reported complaint.The reported issue was unable to be reproduced and the reader functioned as intended.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported her adc blood glucose meter was reading in the incorrect uom and incorrect language.Customer further reported experiencing a loss of consciousness.Customer reported she was unable to self-treated requiring third-party treatment from a healthcare professional of "dextro" and something to eat.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Reader (b)(6) has been returned and investigated.Visual inspection has been performed on the reader and no issues were observed.The unit of measurement of the reader has been tested and no specification was observed.Therefore, issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported her adc blood glucose meter was reading in the incorrect uom and incorrect language.Customer further reported experiencing a loss of consciousness.Customer reported she was unable to self-treated requiring third-party treatment from a healthcare professional of "dextro" and something to eat.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18050707
MDR Text Key327102239
Report Number2954323-2023-48165
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-