MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1TQ290
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for inspection, and the customer's reported issue ¿distal end is broken¿ was confirmed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The customer reported to olympus that the evis lucera elite bronchovideoscope the distal end is broken found during reprocessing.Upon inspection and evaluation, the plastic distal end cover was burnt.There was no report of patient harm or user injury associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during incoming inspection and evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely when the user was performing high-frequency cauterization they came in contact with mucosa and did not push an endo therapy accessory out to the visible position of the green marking in sheath of the accessory, and electrified without contacting electrode of an endo therapy accessory with mucosa.As a result, the distal end was burnt.However, a definitive root cause could not be determined.Instructions for use provides (ifu) the following warnings for high-frequency cauterization of the operation manual.Chapter 4.3 "using endo therapy accessories" warning.Always confirm that the tissue is an appropriate distance away from the distal end of the endoscope.If high-frequency cauterization is performed when the distal end of the endoscope contacts the tissue, patient injury, burns, bleeding, perforation, and equipment damage may occur.Always confirm that the electrode section of the electrosurgical accessory is at an appropriate distance from the distal end of the endoscope.Confirm that the entire green marking (in case of wli observation mode) at the distal tip of the electrosurgical accessory can be observed on the endoscopic image.If the electrode is used when it is too close to the distal end of the endoscope, the endoscope and/or ancillary equipment may be damaged.Patient injury, burns, bleeding, perforation, and/or equipment damage may result.Be sure to contact the electrode section of the high-frequency endo therapy accessory with tissue when performing high-frequency cauterization treatment to prevent equipment damage and operator and/or patient burns." additionally, the instructions for use provides (ifu) provides the following warnings for high-frequency cauterization.To avoid patient injury, burns, bleeding, perforation, and/or damage to the endoscope, never emit laser radiation before confirming that an appropriate distance between the target and the endoscope¿s distal end is maintained and the tip of the laser probe is surely in the correct position in the endoscopic image." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18051681
MDR Text Key327141219
Report Number9610595-2023-16327
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-1TQ290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-290
-
-