SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation.The endoscopy support specialist was on site to perform a radial probe reprocessing and infection control in-service for the staff and observed the person on the attendance sheet performing manual cleaning and preparing the probe for (ethylene oxide) sterilization.The customer was provided infection control information referenced in the user manual and reprocessing manual.The customer used a non-olympus eto sterilization company to sterilize their radial probes.It is recommended that the customer contact the manufacturer of that equipment for proper use.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if the user facility provides any additional information.
 
Event Description
The olympus representative reported that oil leaked from the ultrasonic probe during eto (ethylene oxide) sterilization.The issue occurred during reprocessing.No patient harm was associated with the event.This report is related to the following linked patient identifiers: (b)(6).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18053488
MDR Text Key327153686
Report Number3002808148-2023-12140
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-