Model Number GIF-1200N
Device Problems Device Reprocessing Problem (1091); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/30/2023
Event Type  malfunction  
Manufacturer Narrative
To date, the customer have not returned the device for evaluation.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the scope reprocessed due to overdue water filter replacement used in case.The issue occurred during an unknown event and procedure.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the filter not being replaced was due to the device being reprocessed with the different procedure from the instruction manual.The event can be detected/prevented by following the instructions for use which state: ¿ instructions, operation manual for oer-6 chapter 7 ¿routine maintenance¿ section 7.3 ¿replacing the water filter (maj-2317)¿.The replacement frequency of the water filter was recommended in the instruction manual to be at least once in two months periodically as a guide.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18057243
MDR Text Key327189430
Report Number9610595-2023-16385
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-1200N
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OER-6 ×2.MAJ-2317 ×2.
-
-