Other text: d3, g1, and g2 email is: regulatory.Responses@icumed.H3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection noted it was not possible to detect any issue which could cause the leaking failure mode.Leaking issue was not detected in the leak test.Based on the analysis conducted in the sample provided, air leaking issue was not confirmed in the breathing circuit.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No corrective action is required because the failure mode reported by the customer was not confirmed.
|