Catalog Number C49101320J
Device Problem Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Date 10/10/2023
Event Type  malfunction  
Manufacturer Narrative
Other text: d4: udi and g5: 510k number are unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the anesthesia machine did not pass the leak check test.Possible leak.Patient involvement unknown.
 
Manufacturer Narrative
Other text: d3, g1, and g2 email is: regulatory.Responses@icumed.H3 and h6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection noted it was not possible to detect any issue which could cause the leaking failure mode.Leaking issue was not detected in the leak test.Based on the analysis conducted in the sample provided, air leaking issue was not confirmed in the breathing circuit.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No corrective action is required because the failure mode reported by the customer was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18059079
MDR Text Key327271063
Report Number3012307300-2023-10192
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC49101320J
Device Lot Number4388462
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/17/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-