MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CAPD DISCONNECT Y SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CAPD DISCONNECT Y SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number MRM4366MP
Device Problems Break (1069); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address:(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
T was reported that the capd disposable disconnect y-set was damaged; further described as ¿when the port of the blue ring that goes to the catheter connection was removed, the ring was stuck to the bag." this was observed during use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event, however the patient had been on antibiotics.No additional information is available.
 
Manufacturer Narrative
Additional information: h6 and h10.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPD DISCONNECT Y SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
calle 36 no. 2c-22
apartado aero 2446
cali
CO  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18065961
MDR Text Key327292900
Report Number1416980-2023-05643
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMRM4366MP
Device Lot NumberSE23EH3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2023
Date Device Manufactured06/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-