Catalog Number 2C4041 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a y-type tur/bladder irrigation set was leaking from the drip chamber.This issue was identified during priming.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information was added to d9, h3, h6 and h10.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Pressure and clear passage under water testing was performed and a leak was observed since the drip housing was not sealed in the lower seal.The reported condition was verified.The cause of the condition was due to a improper manual assembly during manufacturing.Should additional relevant information become available, a supplemental report will be submitted.
|
|