Model Number N/A |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that the inserter was taken out of the tray and was found with the threads stripped and coming off.The procedure was completed using another device.There is no additional information available at the time of this report.
|
|
Event Description
|
No further event information is available at the time of this report.
|
|
Manufacturer Narrative
|
Product was returned and evaluated.Medical records were not provided.A visual inspection found that the lead thread was beginning to shear off.The shaft was scratched and exhibited discoloration spots.The strike plate was also dinged and discolored.A review of the device history records identified no deviations or anomalies during manufacturing.The reported event was not related to a combination of products; therefore, a compatibility review was not applicable.The complaint was confirmed via the returned product.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|