MAUDE Adverse Event Report: ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY Back to Search Results
Catalog Number 08056668190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/30/2023
Event Type  malfunction  
Manufacturer Narrative
The serial number of the customer's cobas pro c 503 analytical unit is 2058-03.The investigation is ongoing.
 
Event Description
The initial reporter received questionable tina-quant hemoglobin a1cdx gen.3 result from one patient sample tested on the cobas pro c 503 analytical unit.The initial result was 7.65 %.The repeat result was 5.58%.The repeat result was believed to be the true result based on the patient's history and the glucose result.
 
Manufacturer Narrative
Based on the information provided, the cause of the event could not be determined.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TINA-QUANT HEMOGLOBIN A1CDX GEN. 3
Type of Device
GLYCOSYLATED HEMOGLOBIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18072463
MDR Text Key327378051
Report Number1823260-2023-03520
Device Sequence Number1
Product Code LCP
UDI-Device Identifier07613336120927
UDI-Public07613336120927
Combination Product (y/n)Y
Reporter Country CodeTC
PMA/PMN Number
K072714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08056668190
Device Lot Number70882001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-