Model Number 81000
Device Problems Mechanical Problem (1384); Device Slipped (1584); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2023
Event Type  malfunction  
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No adverse event occurred and no medical intervention was required.The iv pole clamp was installed.
 
Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site and was able to confirm the reported condition.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Correction: a field service engineer adjusted the iv pole release screw.Investigation is in process.A follow-up report will be provided.
 
Manufacturer Narrative
This report is being filed to provide corrected information is provided in investigation: a terumo bct service technician checked out the machine at the customer site and was able to confirm the reported condition.The device serial number history report indicates one further related issue has been reported for this device.One year of service history was reviewed for this device with one further issue related to the reported condition identified (see mdr 1722028-2023-00271) correction: a field service engineer adjusted the iv pole release screw.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related the need for an iv pole screw adjustment.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No adverse event occurred and no medical intervention was required.The iv pole clamp was installed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key18074385
MDR Text Key327417900
Report Number1722028-2023-00361
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK150269
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-